Has the EUDAMED deadline been extended? No. The period for registering devices that were already on the EU market before 28 May 2026 ended at the end of November 2026, and for many manufacturers it ended badly: uploads waited in a queue for a day or more, confirmations came late and the public search slowed to a crawl. An unfinished registration is now a non-compliance. In our reading of the MDR, it is a particular kind of non-compliance, and how you handle the coming weeks decides whether it stays a formality or becomes a problem.
If you need the basics first (who registers, which devices, which data), start with our guide to EUDAMED registration in 2026. This article picks up where that one ends.
Has the EUDAMED deadline been extended?
No, and it cannot be extended informally. The date comes from the regulation itself. Regulation (EU) 2024/1860 amended Article 123 of the MDR so that devices placed on the market before the UDI/device module became mandatory, and still placed on the market afterwards, must be registered within twelve months of the Commission notice of 27 November 2025. The IVDR contains the same rule. Moving that date would take a new amending regulation. An MDCG paper, a Commission statement or a helpdesk reply cannot do it.
The official sources differ on the exact day: the Commission’s timeline shows 28 November 2026, the FAMHP refers to a period ending on 27 November. For the legal analysis below, that difference changes nothing. For your evidence, it matters: if you rely on submissions made in the last days, record the date and time of each one.
Why did so many registrations stall?
Several things happened at once. Many manufacturers left their existing range until the autumn, so a second and larger wave of submissions arrived in the final weeks. Demand for device data downloads rose sharply at the same time, partly from companies checking their own records and partly from automated scraping of the public site. From early November the public site began to refuse automated requests, and that traffic moved to channels that were not built for it. Add routine and emergency maintenance of the underlying infrastructure, and processing times of more than a day became common.
The result is that the word “submitted” now covers three quite different situations. It helps to keep them apart.
The first state is the one that worries people, and the one that needs evidence: you sent the data, but EUDAMED has not yet confirmed whether it accepted it. The second is often misread. For a device covered by a notified body certificate, EUDAMED keeps an accepted registration out of the public site until the notified body links the certificate. Notified bodies have until 28 May 2027 to register certificates issued before mandatory use, so a device can be correctly registered and still be invisible to a buyer who searches for it. Self-declared class I devices do not wait for that step.
What does an unfinished registration mean in law?
Registration is a legal obligation, so a device that should be registered and is not, is non-compliant. Two points narrow the problem considerably.
First, the obligation follows placing on the market, and placing on the market happens unit by unit. The transitional rule covered devices you continue to place on the market. In our reading, units placed on the market before the deadline do not become unlawful because the registration of their UDI-DI is late. The exposure lies in the units you release after it. That is the line to manage, not the stock already in your distributors’ warehouses.
Second, the MDR already has a tool for a device that does not comply but does not put anyone at risk. Under Article 97 MDR, a competent authority that finds such a non-compliance requires the economic operator to end it within a reasonable period that is clearly defined, communicated and proportionate. Only if the operator fails to do so does the authority restrict or withdraw the device. Authorities have used this mechanism before, for legacy devices whose certificates were about to expire, as set out in MDCG 2022-18.
A missing database entry, for a device that is otherwise CE marked, certified and safe, comes close to the textbook case for Article 97. That is our reading, not a formal tolerance. No general grace period has been announced, and each competent authority decides on the facts in front of it. It does mean that the questions an inspector will ask are predictable. When did you start? Can you show it? What has happened since? How quickly did you finish?
What should manufacturers do now?
Treat the late registrations as a deviation in your quality management system, with an owner, a root cause and a closing date. That turns a scramble into a record an authority can follow.
- One list. Put every UDI-DI that is not yet finished on a single list, with its state (in the queue, accepted but not public, or not started) and the date and time of each submission.
- Original evidence. Keep the submission references, the acknowledgement or error messages, screenshots and helpdesk ticket numbers with that list. Export messages as received; a screenshot taken today of last week’s message proves less.
- No duplicates. Do not resubmit records that are already in the queue. Duplicates add load and can create conflicting records, and some device data is hard to change once registered.
- Checks in batches. Check status for groups of devices rather than one at a time, and schedule checks outside peak hours when searches are less likely to time out.
- A decision per device. Decide whether to keep releasing new units while the registration is unfinished. Where the submission is in the queue and documented, carrying on is defensible. Where nothing was submitted, register first, or discuss the situation with your competent authority (in Belgium, the FAMHP).
- A word to your supply chain. Tell importers and main distributors where you stand before they ask. A short letter per product family, listing the UDI-DIs, their state and the expected completion date, prevents blocked orders.
The PRRC is the natural owner of this file. Article 15 MDR does not list registration among the PRRC’s tasks, but a PRRC who confirms the regulatory status of devices before release will want to see it.
What it means for importers and distributors in Belgium
Importers have a specific duty. Under Article 13(4) MDR they verify that the device is registered and add their own details to the registration. An importer that cannot find a device in the public search should not conclude straight away that the manufacturer has failed: for certified devices the record may be accepted but not yet public. Ask the manufacturer for evidence of the registration state from its own view of EUDAMED, keep it on file and set a date to check again.
Distributors are in a different position. They do not register in EUDAMED, and the checks listed in Article 14(2) MDR (CE marking, declaration of conformity, labelling, instructions for use, importer details and UDI) do not include EUDAMED registration as such. Due care still applies, and hospital tenders increasingly ask for registration evidence, so the same request to the manufacturer makes sense. The national registration of distributors with the FAMHP is unchanged; our EUDAMED registration guide covers it.
One reaction to avoid is freezing stock that was lawfully placed on the market before the deadline, only because a public search comes back empty. That creates supply gaps for hospitals without improving compliance.
What comes next?
November was the first test of EUDAMED under full load, not the last. These dates and decisions follow.
- Certificates. Notified bodies have until 28 May 2027 to register certificates issued before mandatory use. Until they do, some registered devices stay out of public view. Ask your notified body for its planning.
- Summaries of safety and (clinical) performance. Under MDCG 2026-4 rev.1, uploading the SS(C)P moves from the notified body to the manufacturer. The upload function is planned for December 2026, and 27 March 2027 is given as the date for devices already on the market.
- Vigilance. The vigilance module will not become mandatory before 2028. Until then, serious incidents and field safety corrective actions go through the national procedures.
- Own-brand ranges. MDCG 2026-5 states that UDI-DIs must be assigned by, and linked to, the legal manufacturer, also when a distributor sells under its own trade name. Belgian distributors with private-label ranges should check their arrangements now.
The lesson from November is simple. A deadline that depends on a shared European system falls, in practice, earlier than the date in the regulation. We plan the next ones with a margin of at least a month. The Commission’s EUDAMED overview and its list of MDCG guidance are the places to watch.
We help manufacturers and authorised representatives build the evidence file, prioritise the open UDI-DIs and write the status letters to their supply chain, and we act as external PRRC under the MDR for micro and small manufacturers. For importers and distributors, we set out what to check and what to keep on file, as part of our regulatory affairs support. Tell us where you stand.
Frequently asked questions
Has the EUDAMED registration deadline been postponed?
No. The deadline is set in the MDR and IVDR as amended by Regulation (EU) 2024/1860, and only a new amending regulation could move it. Unfinished registrations should be completed without delay.
Can I keep supplying a device whose EUDAMED registration is unfinished?
In our reading, units placed on the market before the deadline can continue to be supplied. For new units, the missing registration is a non-compliance. Where the submission was made in time and is documented, an authority would normally expect it to be finished within a short, defined period. Where nothing was submitted, register first or discuss it with your competent authority.
Why can’t I find a registered device in the public EUDAMED search?
Devices covered by a notified body certificate become public only once the notified body links the certificate, which it may do until 28 May 2027. Searches can also time out under heavy load. Ask the manufacturer for evidence from its own EUDAMED view.
What evidence should I keep if EUDAMED delays held up my registration?
Submission references with date and time, the acknowledgement or error messages as received, screenshots and helpdesk ticket numbers, kept with a list of the UDI-DIs concerned and a target completion date.